Packaging Flowchart
Packaging for Infectious substance
UN Packaging Instruction 620
UN Licensed
Watertight Primary
Watertight Secondary
Outer box (Minimal Dimension 100mm)
Absorbent material
Itenised list of contents
Testing Includes
Impact 7kg rod
9M Drop
Stack
95kPa Pressure (on Secondary or primary)
Labels Include
Consignee Address
Shipper Address
Contact Name & Tel. No.
Proper Shipping Name
Hazard Label (Division 6.2)
Inland Post
Shipper Declaration not required
Oveerseas Post
Banned (GBG-05)
Overseas Via Courier
Shipper’s Declaration Required
Diagnostic Packaging
UN Packaging Instruction 650
Watertight Primary
Watertight Secondary
Outer box (Minimal Dimension 100mm for Freight)
Absorbent material
Itenised list of contents
Testing not required but must be capable of passing:-
1.2M Drop
95kPa Pressure (on Secondary or primary)
Labels
Consignee Address
Shipper Address
‘Diagnostic Specimen packed in compliance with IATA Packaging Instruction 650’
Inland or Overseas Post
Inland / Overseas
Couriers